Description
Pt100 HS type – High-temperature standard platinum resistance thermometer
These temperature probes are used for accurate temperature measurements or as reference standards. Each thermometer can be supplied with a calibration certificate issued by a DAkkS or ACCREDIA accredited laboratory and calibrated either by comparison or at fixed points. With fixed-point calibration, the declarable measurement uncertainty ranges from 6 mK to 25 mK up to 660 °C and 150 mK at 850 °C. The typical annual drift is 2.5 mK.
The connection, as well as other parameters, can be configured according to the customer's needs.
Specifications
• Measuring range: 0 °C to 850 °C
• Nominal resistance: Pt100 (100 Ohm)
• Measurement uncertainty: from 6 mK to 150 mK
• Typical annual drift: 2.5 mK
• Connection type: 4-wire
• Connection: as per configuration
• Sheath material: Ceramic with gas
• Inner cable: Platinum
• Dimensions:
– grip: Diam. 20 mm x 100 mm
– diameter: 6mm
– length: according to configuration
– connection cable: according to configuration
• R0: 100 Ohm ± 0.05 Ohm
• Alpha: 0.003850 ± 0.000005
• Self-heating: 4 mK (recommended measuring current 1 mA)
• Sensor tolerance: individual calibration
Options
PT100HS from 0 °C to 850 °C
DAkkS calibration FP – DAkkS calibration at fixed points (up to 660 °C)
DAkkS CF calibration – DAkkS calibration by comparison (up to 850 °C)
ACCREDIA traceable calibration – Comparison calibration issued by an ACCREDIA accredited laboratory for temperature
Made in Europe
Documentation available in the Download area.